TB 9-6625-2117-24
Figure 7. Adjustment and test point location - top view - SN 500000 and above.
15. High Voltage Power Supply
a. Performance Check. Connect digital multimeter between TP962 (fig. 4) for SN
4499999 and below and TP1762 (fig. 5) for SN 500000 and above and ground, using high
voltage probe. If digital voltmeter does not indicate between -3.8612 and -4.0188 V dc,
perform b below.
b. Adjustments. Adjust R963 (fig. 4) for SN 499999 and below and HV ADJ R1765
(fig. 5) for SN 500000 and above for digital multimeter indication of -3.940 V dc (R).