TB 9-6625-2124-35
(1) Connect multimeter to A4TP7 (-) and A4TP11 (+) (fig. 1).
(2) Press CW key and enter data 0 MHz on keyboard. Multimeter will indicate
between -0.004 and +0.004 V dc.
(3) Enter data 5 GHz on keyboard. Multimeter will indicate between 4.996 and
5.004 V dc.
(4) Enter data 10 GHz on keyboard. Multimeter will indicate between 9.996 and
10.004 V dc.
b. Adjustments
BLANK (rear panel).
(3) Press function keys and enter corresponding data on keyboard as listed in (a)
and (b) below:
(a) INSTL PRESET key.
(b) SWEEP TIME key, DATA ENTRY 10 ms.
(5) Press function keys and enter corresponding data on keyboard as listed in (a)
and (b) below:
(a) INSTR PRESET key.
(b) SWEEP TIME key, DATA ENTRY 1 s.
(7) Disconnect frequency counter from TI and connect multimeter to A4TP7 (-) and
7