TB 9-6625-2124-35
(36) Repeat technique of (6) and (7) above for M2 marker. Synthesized signal
generator will indicate between 16.9375 and 17.0625 GHz.
(37) Repeat (21) above.
(38) Synthesized signal generator will indicate between 13.470 and 13.530 GHz.
(39) Repeat (9) above.
(40) Synthesized signal generator will indicate between 17.470 and 17.530 GHz.
b. Adjustments
(1) Connect multimeter to A6TP9 (+) and A6TP10 (-) (fig. 7).
(2) Set switch A3S1 (fig. 9) position 1 to the OPEN (UP) position.
(3) Press INSTR PRESET and CW keys. Enter data 20 GHz on keyboard.
(5) Disconnect multimeter from A6TP9 and connect to A6TP5 (fig. 7). Adjust A6R16
(fig. 7) for multimeter indication -10.000 V dc
(R).
(6) Disconnect multimeter from A6TP5 and connect to A6TP4 (fig. 7). Adjust A6R21
(fig. 7) for multimeter indication of
0.000 V dc
(R).
(7) Disconnect multimeter from A6TP4 and connect to A6TP8 (fig. 7). Adjust A6R34
(fig. 7) for multimeter indication of
-10.000 V dc
(R).
(8) Press CW key and enter data 13.5 GHz on keyboard.
(9) Connect multimeter to A6TP5 (+) and A6TP10 (-) (fig. 7).
30