TB 9-6625-2124-35
(11) Disconnect multimeter from A6TP5 and connect to A6TP8. Adjust A6R24 (fig. 7)
for multimeter indication of 0.000 V dc (R).
(12) Press CW key and enter data 7 GHz on keyboard.
(13) Disconnect multimeter from A6TP8 and connect to A6TP5.
(14) Adjust FREQUENCY VERNIER control for multimeter indication of -3.49675 V dc.
(15) Disconnect multimeter from A6TP5 and connect to A6TP8.
(16) Adjust A6R26 (fig. 7) for multimeter indication of 0.000 V dc.
(17) Press CW key and enter data 2.4 GHz on keyboard.
(18) Repeat (13) above.
(19) Adjust FREQUENCY VERNIER control for multimeter indication of -1.19560 V dc.
(20) Repeat (15) above.
(21) Adjust A6R28 (fig. 7) for multimeter indication of 0.000 V dc (R).
(22) Press CW key and enter data 10 MHz on keyboard.
(23) Repeat (13) above.
(24) Adjust FREQUENCY VERNIER control for multimeter indication of 0.00000 V dc.
(25) Repeat (15) above.
(26) Adjust A6R30 (fig. 7) for multimeter indication of 0.000 V dc (R).
(27) Disconnect multimeter from TI. Connect SWEEP OUTPUT to oscilloscope
CH2 input.
(28) Connect oscilloscope CH1 input to A6TP8. Press INSTR PRESET key.
(a)
CH1 VOLTS/DIV to .5.
(b)
CH2 VOLTS/DIV to .5.
(c)
CH1 input to DC.
(d)
CH2 input to DC.
(30) Set oscilloscope for an X verses Y function operation and adjust oscilloscope
controls as required to obtain a display as shown in figure 10.
31