TB 9-6625-2134-24
(346) Press FREQUENCY RANGE pushbutton to 15 - 21 GHz indication and
center 15 GHz marker on crt with FREQUENCY control.
(347) Press DEGAUSS pushbutton and adjust FREQUENCY control to recenter
signal on crt display.
(348) Adjust PEAKING control to peak 15 GHz signal and adjust FREQUENCY
control to center 19 GHz marker on crt.
(349) Press DEGAUSS pushbutton and adjust R1061 (fig. 25) to peak signal
response on crt.
(350) Adjust FREQUENCY control to center 21 GHz marker indication on crt.
(351) Press DEGAUSS pushbutton and adjust R1063 (fig. 25) to peak signal
response on crt.
(352) Repeat technique of (346) through (351) above for 15, 19, and 21 GHz.
(353) Press FREQUENCY RANGE pushbutton to 5.4 - 18 GHz and center 15
GHz marker on crt with FREQUENCY control.
(354) Press DEGAUSS pushbutton and center 15 GHz marker on crt, using
FREQUENCY control.
(355) Peak 15 GHz marker on crt, using PEAKING control, and press
(356) Center 15 GHz marker on crt, using FREQUENCY control, and press
DEGAUSS pushbutton.
(357) Adjust R1064 (fig. 25) to peak 15 GHz marker on crt if necessary.
(358) Press FREQUENCY RANGE pushbutton to 3.0 - 7.1 GHz and center 5.0
GHz marker on crt using FREQUENCY control.
(359) Press DEGAUSS pushbutton and center 5.0 GHz marker on crt using
FREQUENCY control.
(360) Adjust R1045 (fig. 25) to peak 5.0 GHz marker response on crt.
(361) Press FREQUENCY RANGE pushbutton to 1.7 - 5.5 GHz and center 3.5
GHz marker on crt using FREQUENCY control.
(362) Press DEGAUSS pushbutton and center 3.5 GHz marker on crt, using
FREQUENCY control.
(363) Adjust R1049 (fig. 25) to peak 3.5 GHz signal response on crt.
NOTE
Signal peak for all bands should now track the signal peak
with the PEAKING control index marker near front
panel mark.
44