TB 9-6625-2153-24
Test instrument
Performance specifications
Range: 1 : per step
Accuracy:
r0.25%
9063641
Range: 10 : and above per step
Accuracy:
r0.1%
Decade switch
Decade switch
Min
Max
positions
Ohms
1
1
0.9975000
1.0025000
2
1.9950000
2.0050000
3
2.9925000
3.0075000
4
3.9900000
4.0100000
5
4.9875000
5.0125000
6
5.9850000
6.0150000
7
6.9825000
7.0175000
8
7.9800000
8.0200000
9
8.9775000
9.0225000
10
9.9750000
10.0250000
0
---
---
10
1
9.9900000
10.0100000
2
19.980000
20.020000
3
29.970000
30.030000
4
39.960000
40.040000
5
49.95000
50.05000
6
59.94000
60.06000
7
69.93000
70.07000
8
79.92000
80.08000
9
89.91000
90.09000
10
99.900000
100.10000
0
---
---
100
1
99.90000
100.10000
2
199.80000
200.20000
3
299.70000
300.30000
4
399.60000
400.40000
5
499.50000
500.50000
6
599.40000
600.60000
7
699.30000
700.70000
8
799.20000
800.80000
9
899.10000
900.90000
10
999.00000
1001.00000
0
---
---
Kilohms
1K
1
0.9990000
1.0010000
2
1.9980000
2.0020000
3
2.9970000
3.0030000
4
3.9960000
4.0040000
5