TB 9-6625-2153-24
Table 10. ESI, Models DB-62 and DB-877 - Continued
Test instrument
Decade
Decade switch
DB-62
DB-877
switch
positions
Min
Max
Min
Max
Megohms
1 M:/step
1
0.999800
1.000200
0.999700
1.000300
2
1.999600
2.000400
1.999400
2.000600
3
2.999400
3.000600
2.999100
3.000900
4
3.999200
4.000800
3.998800
4.001200
5
4.999000
5.001000
4.998500
5.001500
6
5.998800
6.001200
5.998200
6.001800
7
6.998600
7.001400
6.997900
7.002100
8
7.998400
8.001600
7.997600
8.002400
9
8.998200
9.001800
8.997300
9.002700
10
9.998000
10.002000
9.997000
10.003000
0
---
---
---
---
1Use appropriate ranges for particular model of TI being tested.
Test instrument
Performance Specifications
Range1
Accuracy
0.01 :/step
r2%
0.1 :/step
r0.4%
1 :/step
r0.1%
1433 Series
10 :/step
r0.04%
100 : and above/step
r0.01%
Decade switch
Decade switch
Min
Max
positions
Ohms
1
00.00980
00.01020
0.01:/step
2
00.01960
00.02040
3
00.02940
00.03060
4
00.03920
00.04080
5
00.04900
00.05100
6
00.05880
00.06120
7
00.06860
00.07140
8
00.07840
00.08160
9
00.08820
00.09180
10
00.09800
00.10200
0
- - -
- - -
1
00.09960
00.10040
0.1:/step
2
00.19920
00.20080
3
00.29880
00.30120
4
00.39840
00.40160
5
00.49800
00.50200
6
00.59760
00.60240
See footnote at end of table.