Table 3. Resistance Decade, APN 9063641 - Continued
Test instrument
Performance specifications
Range: 1 Ω per step
9063641
.25%
Accuracy:
Range: 10 Ω and above per step
.1%
Accuracy:
Ohms
Decade switch
Decade switch
positions
Min
Max
Kilohms - Continued
1K
5
4.9950000
5.0050000
1K
6
5.9940000
6.0060000
1K
7
6.9930000
7.0070000
1K
8
7.9920000
8.0080000
1K
9
8.9910000
9.0090000
1K
10
9.9900000
10.010000
1K
0
---
---
10 K
1
9.9900000
10.010000
10 K
2
19.980000
20.020000
10 K
3
29.970000
30.030000
10 K
4
39.960000
40.040000
10 K
5
49.950000
50.050000
10 K
6
59.940000
60.060000
10 K
7
69.930000
70.070000
10 K
8
79.920000
80.080000
10 K
9
89.910000
90.090000
10 K
10
99.900000
100.10000
10 K
0
---
---
100 K
1
99.900000
100.10000
100 K
2
199.80000
200.20000
100 K
3
299.70000
300.30000
100 K
4
399.60000
400.40000
100 K
5
499.50000
500.50000
100 K
6
599.40000
600.60000
100 K
7
699.30000
700.70000
100 K
8
799.20000
800.80000
100 K
9
899.10000
900.90000
100 K
10
999.0000
1001.0000
100 K
0
---
---
Megohms
1 MEG
1
0.9990000
1.0010000
1 MEG
2
1.9980000
2.0020000
1 MEG
3
2.9970000
3.0030000
1 MEG
4
3.9960000
4.0040000
1 MEG
5
4.9950000
5.0050000
1 MEG
6
5.9940000
6.0060000
6