TB 9-6625-2155-35
Test
instrument
output level
Min
Max
0
-2
2
-10
-12
-8
-20
-22
-18
-30
-32
-28
-40
-42
-38
-50
-52
-48
-60
-62
-58
-70
-73.5
-66.5
-80
-83.5
-76.5
-90
-93.5
-86.5
-100
-103.5
-96.5
(8) Press TI FREQUENCY, 6, and GHz keys. Verify the measuring receiver
indicates within minimum and maximum limits for TI output levels listed in table 10
below, using standard tuned level measurement techniques.
Test
instrument
output level
Min
Max
0
-2
2
-10
-12
-8
-20
-22
-18
-30
-32
-28
-40
-42
-38
-50
-52
-48
-60
-62
-58
-70
-73.5
-66.5
-80
-83.5
-76.5
-90
-93.5
-86.5
-100
-103.5
-96.5
(9) Press TI FREQUENCY, 12, and GHz keys. Verify the measuring receiver
indicates within minimum and maximum limits for TI output levels listed in table 11
below, using standard tuned level measurement techniques.
Test
instrument
output level
Min
Max
0
-2
2
-10
-12
-8
-20
-22
-18
-30
-32
-28
-40
-42
-38
14