TB 9-6625-2164-35
Table 5. Timing Accuracy - Continued
TRIG POS
CURSORS
TIME/DIV
pushbutton
Calibrator
readout
switch settings
indications
Marker Output
Min
Max
s
-95
10
ms
9.99
ms
10.01
ms
100
s
-95
20
ms
19.98
ms
20.02
ms
200
s
-95
50
ms
49.95
ms
50.05
ms
500
1
ms
-95
0.1
s
99.9
ms
100.1
ms
2
ms
-95
0.2
s
199.8
ms
200.2
ms
20
ms1
-95
2
s
1.998
s
2.002
s
500
ms1
-35
2
s
19.98
s
20.02
s
1Press
TRIGGERING/MODE HOLD NEXT pushbutton and wait for display to stop rolling before making cursor
measurement.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.