TB 9-6625-2182-24
Test instrument
DATA
indications
frequency
Min
Max
100 kHz
99.999
kHz
100.001
kHz
500 kHz
499.995
kHz
500.005
kHz
1 MHz
999.990
kHz
1.000010
MHz
5 MHz
4.99995
MHz
5.000050
MHz
10 MHz
9.999900
MHz
10.000100
MHz
50 MHz
49.999500
MHz
50.000500
MHz
100 MHz
99.999000
MHz
100.001000
MHz
Test instrument
DATA
indications
frequency
Min
Max
500 MHz
499.995000
500.005000
1300 MHz
1299.987000
1300.013000
1900 MHz
1899.981000
1900.019000
(6) Press RF OFF/ON pushbutton to OFF.
(7) Disconnect equipment setup.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Set measuring receiver with sensor module to measure RF power with
0.01 db resolution.
NOTE
Zero and calibrate sensor module as necessary.
(2) Connect sensor module to TI OUTPUT RF connector.
(3) Connect TI 10 MHz OUT (rear panel) to measuring receiver external reference input.
(4) Press TI pushbuttons as listed in (a) through (c) below:
(a) INSTR PRESET.
(b) ENTRY - FREQ.
(c) DATA - 30 MHz.
(5) Press TI ENTRY AMPTD pushbutton and enter DATA amplitude for each
row in table 6. Using measuring receiver and RF power measurement techniques,
measured power will indicate within limits specified.