TB 9-6625-2188-35
(8) Adjust FREQUENCY Hz dial and frequency multiplier pushbuttons to values
listed in table 5 below. Differential voltmeter will indicate within limits specified.
(9) Adjust OUTPUT LEVEL (dBm) CAL control fully cw (to detent).
b. Adjustments. Remove left
side cover from TI and
adjust R1423 (fig. 1)
for 0.775 V
Test instrument
FREQUENCY HZ
dial positions
pushbuttons positions (in)
Min
Max
1
x10
0.765730
0.783566
2
x10k
5
x10k
0.756965
0.792639
10
x10k
Figure 1. Output attenuator adjustment -
equipment location (SG505, SG505 opt 01).
6