TB 9-6625-2193-24
Test instrument
Calibrator output
Test instrument indications
Range
Filter
Voltage
Min
Max
Adj. (R)
V
V
V
On
100
mV
50
Hz
---
100.0
95.0
105.0
V
V
V
100
mV
1.0
kHz
---
100.0
95.0
105.0
V
V
V
316
mV
50
Hz
---
316.2
300.2
331.8
V
V
V1
---
316
mV
1.0
kHz
316.2
300.2
331.8
Set calibrator to STANDBY and remove ratio transformer from setup.
1.000 mV
1.0
mV
20
Hz
0.950 mV
1.050 mV
---
1.030 mV1
R452
1.000 mV
1.0
mV
1.0
kHz
0.970 mV
1.000 mV
Off
1.0
mV
50
kHz
0.970 mV
1.030 mV
---
3.162 mV
On
3.16 mV
20
Hz
3.002 mV
3.318 mV
---
3.255 mV1
R583
3.162 mV
3.16 mV
1.0
kHz
3.065 mV
3.162 mV
Off
3.16 mV
50
kHz
3.065 mV
3.255 mV
---
10.00
mV
On
10.0
mV
20
Hz
9.50
mV
10.50 mV
---
10.30 mV1
10.00
mV
10.0
mV
1.0
kHz
9.70
mV
---
10.00
mV
Off
10.0
mV
50
kHz
9.70
mV
10.30 mV
---
31.62
mV
On
31.6
mV
20
Hz
30.02
mV
33.18 mV
---
32.55 mV1
R383
31.62
mV
31.6
mV
1.0
kHz
30.65
mV
31.62
mV
Off
31.6
mV
50
kHz
30.65
mV
32.55 mV
---
100.0
mV
On
100
mV
20
Hz
95.0
mV
105.0
mV
---
mV1
100.0
mV
100
mV
1.0
kHz
97.0
mV
103.0
---
100.0
mV
Off
100
mV
50
kHz
97.0
mV
103.0
mV
---
C113
100.0
mV
100
mV
100
kHz
95.0
mV
105.0
mV
316.2
mV
On
316
mV
20
Hz
300.2
mV
331.8
mV
---
mV1
R273
316.2
mV
316
mV
1.0
kHz
306.5
mV
325.5
316.2
mV
Off
316
mV
50
kHz
306.5
mV
325.5
mV
---
316.2
mV
316
mV
500
kHz
300.2
mV
331.8
mV
---
1.000 V
On
1.0
V
20
Hz
0.950 V
1.050 V
---
1.030 V1
1.000 V
1.0
V
1.0
kHz
0.970 V
---
1.000 V
Off
1.0
V
50
kHz
0.970 V
1.030 V
---
1.000 V
1.0
V
500
kHz
0.950 V
1.050 V
---
3.162 V
On
3.16 V
20
Hz
3.002 V
3.318 V
---
3.255 V1
3.162 V
3.16 V
1.0
kHz
3.065 V
---
3.162 V
Off
3.16 V
50
kHz
3.065 V
3.255 V
---
Press the following keys: /9//9//.//1//SHlFT//SF/
C53, 4
3.162 V
3.16
V
100
kHz
3.002 V
3.318 V
Press the following keys: /0//SHIFT//SF/
C63, 4
3.162 V
3.16
V
100
kHz
3.002 V
3.318 V
10.00
V
On
10
V
20
Hz
9.50
V
10.50
V
---
10.00
V
10
V
1.0
kHz
9.70
V
10.30
V
---
10.00
V
Off
10
V
50
kHz
9.70
V
10.30
V
---
10.00
V
10
V
500
kHz
9.50
V
10.50
V
---
10.00
V
10
V
1.0
MHz
9.00
V
11.00
V
---
See footnotes at end of table.
CHANGE 1