TB 9-6625-2193-24
Table 3. Voltage Accuracy - Continued
Test instrument
Calibrator output
Test instrument indications
Range
Filter
Voltage
Min
Max
Adj. (R)
31.62 V
On
30
V
20
Hz
28.50 V
31.50 V
---
31.62
V
30
V
1.0
kHz
29.10
V
30.90
V
---
31.62
V
Off
30
V
50
kHz
29.10
V
30.90
V
---
31.62
V
30
V
500
kHz
28.50
V
31.50
V
---
31.62
V
20
V
1.0
MHz
18.00
V
22.00
V
---
100.0
V
On
100
V
20
Hz
95.0
V
105.0
V
---
100.0
V
100
V
1.0
kHz
97.0
V
103.0
V
---
100.0
V
Off
100
V
50
kHz
97.0
V
103.0
V
---
100.0
V
40
V
500
kHz
38.0
V
42.0
V
---
RESET calibrator and connect power amplifier to setup.
316.2
V
On
300
V
40
Hz
285.0
V
315.0
V
---
316.2
V
300
V
1.0
kHz
291.0
V
309.0
V
---
316.2
V
Off
300
V
50
kHz
291.0
V
309.0
V
---
316.2
V
300
V
100
kHz
285.0
V
315.0
V
---
RESET calibrator and disconnect power amplifier from setup.
1Record
TI indications.
4C5 and C6 interact. Repeat if C6 is adjusted.