TB 9-6625-2210-35
(7) TRANSMIT TLP/ABS to ABSOLUTE.
e. Press Pg Dn key and set parameters as listed in (1) through (10) below:
(1) SF SKIP to ON.
(2) INCR RATE to 3.0.
(3) SWEEP to OFF.
(4) LO FREQ (Band g (4) below must be VOICE) to 20 Hz.
(5) HI FREQ to 200 Hz.
(6) FREQ INCR to 10 Hz.
(7) Y-LO to -01.0.
(8) Y-HI to +01.0.
(9) LIMITS to C4.
(10) PLOT to OFF.
f. Press Pg Dn key and set parameters as listed in (1) through (4) below:
(1) TRANSMIT LEVEL to 00.0 dBm.
(2) TRANSMIT FREQ to 1000 Hz.
(3) 60 Hz FILTER to OUT.
(4) BAND to VOICE.
g. Allow 5 minutes for warm-up before beginning test.
a. Performance Check
(1) Remove disk drive cover by removing two screws, top half of cover, turning lever
to vertical, and pulling remainder of cover directly forward.
(2) Insert calibration disk into slot with label to left and return lever to horizontal
position.
(3) Press the following keys:
(a) HOME.
(b) HOME.
(c) F10.
(d) A.
(e) :.
(f) RETURN.