TB 9-6625-2210-35
(8) Type WBEDDCAL, press return key, and follow instructions on TI screen.
(9) The test set will display the following:
"Writing calibration data to EEPROM. Calibration sequences completed.
Please write protect EEPROM. Dump calibration data? (y/n):"
(a) Type the following: N
(b) The test set will display the following: A:\>
(10) Remove calibration disk and replace in protective jacket.
(11) Switch power off. Replace cover and disk drive cover.
b. Adjustments. No further adjustments can be made.
a. Performance Check
(1) Switch TI power on and wait for main menu to appear on screen.
(2) Connect TI RCV to XMT 2W/4W using type 310 phone plug and cable.
(3) Select F2: TIME/RL from main menu and H:
SELF
TEST
from
TRANSMISSION MEASUREMENT FUNCTIONS menu.
(4) Select A: P/AR from SELF TEST menu and press Run/Halt key when P/AR
SELF TEST menu appears.
(5) After PASS appears in both right hand boxes, press Run/Halt then Home keys.
(6) Select B: ENVELOPE DELAY and press Run/Halt key when ENVELOPE
DELAY SELF TEST screen appears.
(7) After PASS appears in three boxes on right of screen, press Run/Halt then
Home keys.
when INTERMODULATION SELF TEST screen appears.
(9) After PASS appears in the three boxes on right of screen, press Run/Halt then
F1 keys.
(10) Disconnect cable from TI.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect calibrator to TI RCV connector using a 310 phone plug and cable.
(2) Select F2: TIMS/RL from main menu and B: LEVEL/FREQ/ENV DELAY from
TRANSMISSION MEASUREMENT FUNCTIONS menu.
(3) Press Pg Dn key two times and adjust calibrator for 2.313 V and 1000 Hz.
(4) Press Run/Halt key. TI will display 9.4 to 9.6 dBm and 999 to 1001 Hz.
CHANGE 1