TB 9-6625-2213-24
Table 2. Minimum Specifications of Equipment Required - Continued
Manufacturer and model
Common name
Minimum use specifications
(part number)
LOW PASS FILTER
Telonic. Model TLC200-6EF
NO. 5
Insertion loss: <1.0 dB
(TLC200-6EF)
Stopband attenuation:
250 to 255 MHz: 40 dB
255 to 400 MHz: 42 dB
LOW PASS FILTER
Telonic. Model TLC316-6EF1
NO. 6
Insertion loss: <1.0 dB
(TLC316-6EF1)
Stopband attenuation:
400 to 632 MHz: 42 dB
Telonic, Model TLC450-6EF
LOW PASS FILTER
Cut off frequency: 450 MHz
(TLC450-6EF)
NO. 7
Insertion loss: <1.0 dB
LOW PASS FILTER
Cut off frequency: 700 MHz
Telonic, Model TLC700-6EF1
NO. 8
Insertion loss: <1.0 dB
(TLC700-6EF1)
Telonic, Model TLS1225-5EF1
LOW PASS FILTER
Cut off frequency: 1225 MHz
(TLS1225-5EF1)
NO. 9
Insertion loss: <1.0 dB
Stopband Attenuation:
1400 to 1700 MHz: 10 dB
1700 to 2450 MHz: 42 dB
MEASURING
Frequency range: 1 to 1000 MHz
Measuring receiver system
RECEIVER
Measurement range: +13 to -70.5 dBm
N5530S consisting of: Spectrum
Analyzer Agilent, Model E4440A
Accuracy: 0.025 dB/10 dB
(E4440A), Power Meter, Agilent,
Model E4419B (E4419B), and
Sensor Module Agilent, Model
N5532A opt. 518 (518)
Dc mode:
Hewlett-Packard, Model 3458A
Range: -20.2 to +20.2
(3458A)
Accuracy: 25%
Ac mode: Able to measure less than 1 mV rms
Able to measure 5 mV p-p
Agilent, OS-303/G (OS-303/G)
POWER SPLITTER
Frequency range: 10 to 1000 MHz
Weinschel, Model 1870A
(7916839)
Port-to-port
tracking accuracy:
10 MHz to 1 GHz
0.15 dB
Frequency range: 1 to 1000 MHz
Aeroflex, Model 2023B (2023B) or
Power output range: +13 to -66.9 dBm
(SG-1207/U)
Power accuracy of measuring receiver
SYNTHESIZER/LEVEL
Frequency range: 20 kHz
Hewlett-Packard Model
3335AOPT 001-K06 (MIS-35938)
Weinschel, Model AF117A-69-34
VARIABLE
Frequency range: 30 and 100 MHz
(AF117A-69-34)
ATTENUATOR
Attenuation range: 0 to 10 dB
4