TB 9-6625-2213-24
a. Performance Check
-6.99 dBm.
(2) Connect equipment as shown in figure 5, CONNECTION A.
frequency counter.
(4) Set AMPLITUDE CHANNEL switch to A and AMPLITUDE mV-RANGE-DB
switch to 300 mV.
(5) Adjust FREQ RANGE-MHz control so setting includes 400 MHz.
(6) Adjust signal generator amplitude controls for a TI AMPLITUDE meter
indication of 100 mV.
(7) Set PHASE RANGE switch to 180 and PHASE METER OFFSET switch to
0 degrees.