TB 9-6625-2214-24
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286; AN/GSM-287; or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Range: 105 to 125 V ac
Ridge, Model 9020A (9020A)
Accuracy: 1%
CALIBRATOR
Range: -400 to 400 V dc
Fluke, Model 5720A (5720A) (p/o
Accuracy: 0.016%
MIS-35947)
Resolution: 1 part in 10-10
Tracor, Model 527E (527E)
METER
Agilent, Model 33250A (33250A)
Range: 10 Hz to 10 MHz
Amplitude: 0 to 100 mV
Range: 2.95 to 15.9 V dc
Hewlett-Packard, Model 3458A
Accuracy: 0.3%
(3458A)
X-Y mode capability
Agilent, OS-303/G (OS-303/G)
Aeroflex, Model 2023B (2023B)
Range: 10 to 500 MHz
Amplitude: 0 to 25 mV
or (SG-1207/U)
Datum, Model ET6000-75
TIME/FREQUENCY
-10
(13589305)
WORKSTATION
Accuracy: 1 part in 10
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.