TB 9-6625-2215-24
(b) ATTEN switch to X1.
(c) SENSITIVITY control fully cw.
(e) GATE TIME SEC switch to 10.
(2) Connect function generator output to TI 1 MΩ input, using termination.
(3) Adjust function generator for a 10 Hz sine wave frequency signal and amplitude
to minimum output.
(4) Slowly increase function generator amplitude until measuring system indicates
a stable count of the applied frequency. Function generator output will be within limits
(5) Repeat technique of (3) and (4) above for function generator frequencies listed in
table 6. Function generator adjusted output will be within limit specified in table 6.
Function generator output
Frequency
Amplitude limits
(Hz)
(mV)
≤ 100
10
≤ 50
50
≤ 25
100
≤ 25
1000
≤ 25
10000
≤ 25
1
100000
≤ 25
2
1000000
≤ 100
500000003
Set TI GATE TIME SEC switch to 1.
1
Set TI GATE TIME SEC switch to .1.
2
3 Set TI RANGE MHz switch to 50 (blue).
(6) Set RANGE MHz switch to 50 (white) and GATE TIME SEC switch to 10.
(7) Adjust function generator frequency for 10 Hz and amplitude to minimum
output.
(8) Slowly increase function generator amplitude until measuring system indicates
a stable count of the applied frequency. If function generator output exceeds limits
specified in first row of table 7, perform b below.
(9) Repeat technique of (7) and (8) above at function generator frequencies listed in
12