TB 9-6625-2215-24
Test instrument indications
Test instrument
Calibrator output
(Vac)
RANGE
pushbuttons
Voltage
Min
Max
10
9.000
40
8.9098
9.0902
10
9.000
1000
8.9098
9.0902
10
9.000
100000
8.9018
9.0982
100
90.000
40
88.600
91.400
100
90.000
100
88.600
91.400
100
90.000
500
88.600
91.400
1000
900.00
500
886.00
914.00
1000
900.00
100
886.00
914.00
1000
900.00
40
886.00
914.00
b. Adjustments
(1) Adjust A2 R62 (fig. 3) to obtain best in tolerance compromise of error on
each range.
(2) Adjust A2 R51 (fig. 3) to obtain best in tolerance compromise of error on
each range.
a. Performance Check
(1) Press FUNCTION OHM and RANGE 10 kΩ pushbuttons.
(2) Connect calibrator to TI HI and LO terminals.
(3) Set calibrator output to 9.9000 kΩ. TI will indicate within limits specified in
(4) Repeat technique of (1) and (3) above using TI range settings and calibrator
Calibrator
Test instrument
Output
Limits
RANGE
(Ω)
Min
Max
9.8502 kΩ
9.9498 kΩ
10 kΩ
9.9000 k
98.502 kΩ
99.498 kΩ
100 kΩ
99.000 k
9.8258 MΩ
9.9742 MΩ
10 MΩ
9.9000 M
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment and reinstall protective cover on TI.
b. Annotate and affix DA label/form in accordance with TB 750-25.