TB 9-6625-2232-35
(2) Press TI RESET key then FUNCTION ~V key.
(3) Set calibrator for a 1 V, 1 kHz output.
(4) Press TI RANGE ⇑ or ⇓ key to 100 V range. Adjust K until TI indicates
01.000 1 count (R).
(5) Press TI RANGE ⇓ key to 1 V range. Adjust L until TI indicates 1.00000 3
counts (R).
(6) Press TI RANGE ⇑ key to 1 kV range. Adjust either M or N until TI indicates
001.00 2 counts (R).
(7) Repeat (4) through (6) above until no further adjustments are necessary.
(8) Set calibrator to STANDBY and disconnect from TI.
(9) Short TI VOLTS/2WRΩ /4WRΩ HI and LO.
(10) Press FUNCTION ∼ + = key and RANGE ⇑ or ⇓ key to 1 V range. Adjust N
for a minimum TI indication.
(11) Press FUNCTION ~V key and perform (a) through (c) below:
(a) Record TI indication.
(b) Adjust M for a minimum TI indication.
(c) Continue adjusting M in the same direction until TI indication is within
2 counts of indication recorded in (a) above.
(12) Repeat (10) and (11) above until both indications are within 2 counts of each
other.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.