TB 9-6625-2233-24
Test instrument
Performance specifications
parameters
Sine wave
Range: 100 mV to 1 V
amplitude6 (cont)
Accuracy: (30% +2 mV)
Range: 1 to 100 mV
Accuracy: (30% +1 mV)
Square/triangle
Range: 2 mHz to 99.99 kHz
amplitude
Range: 10 to 30 V
(all values p-p)6
Accuracy: (3% +20 mV)
Range: 1 to 9.99 V
Accuracy: (3% +l0 mV)
Range: 0.1 to 0.999V
Accuracy: (3% +4 mV)
Range: 10 to 99.9 mV
Accuracy: (3% +3 mV)
Range: 1 to 9.99 mV
Accuracy: (3% +2 mV)
Range: 100 to 999.9 kHz
Range: 10 to 30 V5
Accuracy: (5% +20 mV)
Range: 1 to 9.99 V
Accuracy: (5% +10 mV)
Range: 0. 1 to .999 V
Accuracy: (5% +4 mV)
Range: 10 to 99.9 mV
Accuracy: (5% +3 mV)
Range: 1 to 9.99 mV5
Accuracy: (5% +2 mV)
Range: 1 to 4.999 MHz
Range: 10 to 30 V
Accuracy: (6% +20 mV)
Range: 1 to 9.99 V
Accuracy: (6% +10 mV)
Range: 0.1 to 0.999 V
Accuracy: (6% +4 mV)
Range: 10 to 99.9 mV
Accuracy: (6% +3 mV)
Range: 1 to 9.99 mV5
Accuracy: (6% +2 mV)
Range: 5 to 19.99 MHz (square wave only)
Range: 10 to 30 V
Accuracy: (12% +20 mV)
Range: 1 to 9.99 V
Accuracy: (12% +10 mV)
Range: 0.1 to .999 V
Accuracy: (12% +4 mV)
See footnotes at end of table.