Table 10. Triangle Wave Amplitude Accuracy 100 kHz to 4.999 MHz
Table 13. Square Wave Amplitude 100 kHz to 4.999 MHz

TB-9-6625-2233-35 Function Generator SG-1288/G Manual
Page Navigation
  16    17    18    19    20  21  22    23    24    25    26  
TB 9-6625-2233-35
(16) Repeat technique of (13) through (15) above substituting settings and
indications listed in table 11.
Table 11. Triangle Wave Amplitude 1 to 4.999 MHz
Oscilloscope indications
Frequency
Output
(V p-p)
(Hz)
(V p-p)
Min
Max
1
M
9.99
9.38
10.6
1
M
10
9.38
10.6
1
M
15
14.1
15.9
4.999  M
9.99
9.38
10.6
4.999  M
10
9.38
10.6
4.999  M
15
14.1
15.9
b. Adjustments. No adjustments can be made.
15. Square Wave Amplitude
a. Performance Check
(1) Connect UNBALANCED output to multimeter input using 50 Ω feedthrough termination.
(2) Press pushbuttons as listed in (a) through (i) below:
(a)
Calibrate (allow enough time to complete and verify AUTOCALIBRATED
is displayed).
(b)
RESET.
(c)
FUNCTION (square wave).
(d)
AMPLITUDE.
(e)
.009.
(f)
ENTER.
(g)
FREQUENCY.
(h)
50.
(i)
ENTER.
(3) Multimeter will indicate between 3.37 and 5.64 mV rms.
(4) Repeat technique of (2) (c) through (i), and (3) above substituting settings and
indications listed in table 12.
Table 12. Square Wave Amplitude Accuracy 50 Hz to 99.99 kHz
Test instrument
Multimeter
Frequency
Output
rms indications
(Hz)
(V p-p)
Min
Max
50
.01
3.35
mV
6.65
mV
50
.099
46.5
mV
52.5
mV
50
.1
46.5
mV
53.5
mV
50
.999
.483  V
.516  V
50
1
.480  V
.520  V
50
9.99
4.84
V
5.15
V
50
10
4.84
V
5.16
V
50
15
7.27
V
7.74
V
99.9  k
.01
3.35
mV
6.65
mV