TB 9-6625-2239-35
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Voltage output:
Fluke, Model 5820A-5C-GHZ
CALIBRATOR
Range:
30 mV to 30 V
(5820A-5C-GHZ)
Accuracy: 0.25%
Time markers:
Range:
1 ns to .5 s
Accuracy:
.5%
Risetime:
.875 ns (1 ns or less)
Aberrations:
.875% (l %)
Level sine wave:
Amplitude: 30 mV
Capability: >100 MHz
DIGITAL
Range: 0 to +55 V
Fluke, Model 8840A/AF05
Accuracy: .25%
(AN/GSM-64D)
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d.
When indications specified in
paragraphs 8 through 12 are not within
tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,
repeat paragraphs 8 through 12. Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.