TB 9-6625-2239-35
(6) Press X10 MAG pushbutton to on (in).
(7) Set oscilloscope calibrator MARKER output to .1 ms.
(8) Repeat technique of (4) and (5) above. If oscilloscope calibrator err display
readout does not indicate within 3%, perform b (l) below.
(9) Press and release X10 MAG pushbutton to off (out).
NOTE
Perform (10) through (12) below for Tektronix, Types 465 and
465B only. Perform (13) through (26) below for all models.
(10) Set A AND B TIME/DIV switches to 10 s.
(11) Set oscilloscope calibrator MARKER output to 10 s.
(12) Repeat technique of (4) and (5) above. If oscilloscope calibrator err display
readout does not indicate within 2%, perform b (l) below.
(13) Set A AND B TIME/DIV switches to 0.5 s.
(14) Set oscilloscope calibrator MARKER output to .5 s.
(15) Repeat technique of (4) and (5) above. If oscilloscope err display readout does
not indicate within 2%, perform b (l) below.
(16) Set A AND B TIME/DIV switches to .05 s and press X10 MAG pushbutton to
on (in).
(17) Set oscilloscope calibrator MARKER output to 5 ns.
(18) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton for 1
sine wave per division excluding first ten divisions displayed on TI. If oscilloscope
calibrator err display readout does not indicate within 3%, perform b (l) below.
(19) Repeat technique of (1) and (3) through (9) above for each TI A AND B
TIME/DIV switch and
oscilloscope calibrator MARKER output
setting listed in
table 5.
Oscilloscope calibrator err display readout will indicate within 2% with X10 MAG
pushbutton pressed to off (out) and 3% with Xl0 MAG pushbutton pressed to on (in).
(20) Press and release X10 MAG pushbutton to off (out).
(21) Press HORIZ DISPLAY B (DLY'D) pushbutton to on (in).
(22) Set A TIME/DIV switch to 2 ms and B TIME/DIV switch to 1 ms.
(23) Repeat technique of (3) through (5) above. If oscilloscope calibrator err display
readout does not indicate within 2%, perform b (2) below.
(24) Set A TIME/DIV switch to 1 s and B TIME/DIV switch to 0.5 s.
Set
oscilloscope calibrator MARKER output to .5 s.
(25) Repeat technique of (4) and (5) above. If oscilloscope calibrator err display
readout does not indicate within 2%, perform b (2) below.
NOTE
A TIME/DIV switch must not be set to the same setting as the
B TIME/DIV switch to perform (26) below.
11