TB 9-6625-2240-35
(3) Press HORIZ DISPLAY B DLY'D pushbutton (in).
(4) Press X10 MAG (IN) pushbutton to (out).
(5) Set TI VOLTS/DIV, TIME/DIV switch settings and oscilloscope calibrator
(6) Adjust horizontal POSITION control to align start of sweep with center vertical
graticule line. Leading edge of displayed time marker will be aligned with the center
vertical graticule line; if not, perform adjustment listed in table 10.
Table 10. Delay Time and Differential Measurement
Test
Oscilloscope
instrument
Test instrument
VOLTS/DIV
TIME/DIV
calibrator
switch
switch settings
Test instrument
MARKER output
settings
adjustment
settings
A
B
.5
b(2)
10 s
0.1 s
10 s
(7) Set TI VOLTS/DIV, TIME/DIV switch and oscilloscope calibrator output as
(8) Rotate DELAY TIME POSITION dial to 9.00. Leading edge of displayed time
marker will be aligned with center vertical graticule line; if not, perform adjustment listed
Table 11. Delay Time and Differential Measurement
Test instrument
DELAY TIME
Test
POSITION
Test
Oscilloscope
instrument
Test instrument
dial limits
VOLTS/DIV
TIME/DIV
instrument
calibrator
()
MARKER
switch
switch settings
adjustment
output settings
settings
A
B
.05
.5
10 s
0.1 s
10 s
b(3)
(9) Rotate DELAY TIME POSITION dial to 1.00.
(10) Set TI VOLTS/DIV, TIME/DIV switch settings and oscilloscope calibrator
(11) Adjust horizontal POSITION control to align 1st marker with center vertical
graticule line using test instrument settings and oscilloscope calibrator MARKER output
(12) Rotate DELAY TIME POSITION dial to align 9th time marker with center
vertical graticule line. DELAY TIME POSITION dial will indicate as listed in table 12. If
not, perform test instrument adjustments listed in table 12.
13