TB 9-6625-2246-40
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Fluke, Model 5720A (5720A)
CALIBRATOR
Dc voltage
(p/o MIS-35947); w amplifier,
Range: 10 to 1000 V
Fluke 5725A/AR (5725A/AR)
Accuracy:1
Ac voltage:
Voltage: 10 mV to 700 V
Accuracy: (%)
Frequency (kHz)
Volts
0.015
0.030
0.090
15
45
90
200
450
950
10
mV
0.18
0.1
0.077
0.07
0.1
0.26
1.17
---
---
100
mV
---
---
---
---
---
0.17
0.62
.9
1.7
1
V
---
---
0.017
---
0.047
---
---
---
---
10
V
0.1
0.042
---
---
---
---
---
---
---
20
V
---
---
---
---
---
---
1.125
1.5
3.75
100
V
0.1
0.042
0.017
0.01
0.047
0.14
---
---
---
700
V
---
---
0.027
0.022
0.061
---
---
---
---
Wideband ac voltage:
Voltage: 100 mV to 3 V
Accuracy: 3.75%
Dc current:
Range: 100 A to 1A
DC REFERENCE
Range: 10 V
Fluke, Model 732A (732A)
STANDARD
Accuracy: Test report
DC VOLTAGE DIVIDER
ESI, Model RV726 (MIS-10295)
Range: 10:1
Accuracy:1
FUNCTION/ARBITRARY
Range: 20 Hz, 10 MHz 1 V rms
Agilent, Model 33250A
(33250A)
Accuracy: 0.0025%
Agilent, Model 3458A (3458A)
Range: 1 V dc
Accuracy:2
Resolution: 1 V
NULL METER
Fluke, Model 845AB (845AB)
A
ESI, Model SP 2980 (MIS-
Range: 1, 10, and 100 M
10281)
MEASURING SYSTEM
ccuracy:3, 4
A
Beckman, Model CR10M
Range: 1 and 10 M
(8598965)
ccuracy:3
NO. 1
See footnotes at end of table.
5