TB 9-6625-2246-50
Table 7. Ac Voltage Continued
Test
instrument
Calibrator output settings
Test instrument indications
range settings
Voltage
Min
Max
100
V
100
V
15
Hz
99.5800 V
100.4200 V
100
V
100
V
30
Hz
99.8300 V
100.1700 V
100
V
100
V
90
Hz
99.9300 V
100.0700 V
100
V
100
V
15
kHz
99.9600 V
100.0400 V
100
V
100
V
45
kHz
99.8100 V
100.1900 V
100
V
100
V
90
kHz
99.4200 V
100.5800 V
100
V
20
V
200
kHz
19.1000 V
20.9000 V
100
V
20
V
450
kHz
18.8000 V
21.2000 V
100
V
20
V
950
kHz
17.0000 V
23.0000 V
1000
V
700
V
90
Hz
699.240 V
700.760 V
1000
V
700
V
15
kHz
699.380 V
700.620 V
1000
V
700
V
45
kHz
698.550 V
701.450 V
(6) Press calibrator RESET key and disconnect from TI.
(7) Connect calibrator WIDEBAND to TI Input HI and LO using cable and
termination supplied with calibrator.
(8) Press TI Guard Open/To LO pushbutton to To LO.
(9) Press FUNCTION/RANGE Range ⇓ key to 100 mV range.
(10) Set calibrator for a 100 mV, 2 MHz wideband output. TI will indicate between
85 and 115 mV.
(11) Press FUNCTION/RANGE Range ⇑ key to 1 V range.
(12) Set calibrator for a 1 V, 2 MHz wideband output. TI will indicate between 0.85
and 1.15 V.
(13) Press FUNCTION/RANGE Range ⇑ key to 10 V range.
(14) Set calibrator for a 3 V, 2 MHz wideband output. TI will indicate between 2.2
and 3.8 V.
b. Adjustments. If an
out-of-tolerance condition exists, perform paragraph 18
below and repeat 11 above.
a. Performance Check
(1) Press keys as listed in (a) through (f) below:
(a) FUNCTION/RANGE blue shift key then Reset/⇒ key.
(b) FUNCTION/RANGE FREQ .
(c) FUNCTION/RANGE blue shift then MENU E/NPLC.