TB 9-6625-2249-24
(10) Configure measuring receiver to measure frequency at 2 GHz.
(11) Configure measuring receiver to measure power.
Indication will be equal to or
greater than 20 dB.
(13) Configure measuring receiver to measure frequency. Indication will be the same
as frequency of the synthesized signal generator.
(14) Configure measuring receiver to measure power. Indication will be equal to or
greater than 20 dB.
(15) Repeat technique of (9) through (14) above for remaining test frequencies listed
b. Adjustments. No adjustments can be made.
9. Leveled Output Power Gain and flatness
a. Performance Check
(2) Set synthesized signal generator for a 2 GHz output signal at +5dBm.
(3) Adjust synthesized signal generator level output for a +19 dBm indication on TI
POWER LEVEL meter.
(4) Connect microwave frequency counter to point A in figure 3.
Figure 3. Leveled out output power gain and flatness - equipment setup.
(5) Adjust synthesized signal generator for a 2.0 GHz indication on microwave
frequency counter.
(6) Disconnect microwave frequency counter from point A of figure 3.
(7) Connect power meter to point A of figure 3.