SECTION III. CALIBRATION PROCESS
Calibrator Output Accuracy Test

Calibration Procedure For Spectrum Analyzer An/Usm-489A Page Navigation
  1    2  3  4    5    6    7    8  
img
. . . .
TB 9-6625-2250-24
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.
NOTE
Verify the proper CAL FACTORS are loaded for the
measuring receiver's sensor module.
a. Connect TI to 115 V ac power source.
b. Press LINE pushbutton to on and allow at least 30 minutes for TI to warm-up
and stabilize.
NOTE
1. [ ] Denotes softkey on display.
2. Refer to C in figure 1 for softkey location.
3. [ ] Softkeys and keys will be referred to as keys in this
procedure.
Figure 1. Crt adjustment pattern.