TB 9-6625-2250-24
MKR indication will be between 3.8 and 4.2 GHz.
(57) TI
a. Performance Check
NOTE
Set the measuring receiver to measure power.
Zero and
calibrate measuring receiver sensor module.
Figure 5. Frequency response - equipment setup.
(2) Press TI keys and enter values using DATA keys as listed in (a) through (g) below:
INSTRUMENT STATE PRESET.
(a)
INSTRUMENT STATE RECALL.
(b)
[MORE].
(c)
[FACTORY PRSEL PK].
(d)
(e)
SPAN then [SPAN WIDTH] to 1 MHz.
(f)
AMPLITUDE then [LOG dB/DIV] to 2 dB/DIV.
(g)
(4) Press MARKER PEAK SEARCH key.
(5) Set measuring receiver to measure RF power in dBm at 300 MHz.
(6) Adjust synthesized signal generator level output for a TI MKR amplitude
indication between -9.95 and -10.05 dBm.
33