Figure 5. Frequency response - equipment setup.
Figure 6. Frequency response - equipment setup.

Calibration Procedure For Spectrum Analyzer An/Usm-489A Page Navigation
  26    27    28    29    30  31  32    33    34    35    36  
img
. . .
TB 9-6625-2250-24
(7) Record measuring receiver indication.
(8) Set synthesized signal generator frequency to the first frequency listed in table 16.
(9) Set measuring receiver to measure RF power in dBm at synthesized signal
generator frequency.
(10) Press TI FREQUENCY key then [CENTER FREQ] to 18 GHz.
(11) Press MARKER PEAK SEARCH key.
(12) Repeat (6) above. Measuring receiver indication will be within 6.1 dB of
indication recorded in (7) above.
(13) Repeat technique of (8) through (12) above for the remaining synthesized signal
generator frequency and TI [CENTER FREQ] settings listed in table 16.
Table 16. Frequency Response
Synthesized signal generator frequency
and test instrument
[CENTER FREQ]
settings
18.0
GHz
17.5
GHz
17.0
GHz
16.5
GHz
16.0
GHz
15.5
GHz
15.0
GHz
14.5
GHz
14.0
GHz
13.5
GHz
13.0
GHz
12.5
GHz
12.0
GHz
11.5
GHz
11.0
GHz
10.5
GHz
10.0
GHz
9.5
GHz
9.0
GHz
8.5
GHz
8.0
GHz
7.5
GHz
7.0
GHz
6.5
GHz
6.0
GHz
5.5
GHz
5.0
GHz
4.5
GHz
4.0
GHz
3.5
GHz
3.0
GHz
2.5
GHz