TB 9-6625-2251-35
Table 4. Channel 2 Cursor Volts Accuracy - Continued
Range
Indications
VOLTS/DIV
settings
Min
Max
100 mV
594 mV
606 mV
200 mV
1.188
V
1.212
V
500 mV
2.970
V
3.030
V
1
V
5.94
V
6.06
V
2
V
11.88
V
12.12
V
5
V
29.7 V
30.3 V
i. Select: CURSOR FUNCTION and set: Time to on. Adjust TI HORIZONTAL A
AND B SEC/DIV controls for 500 s indication on crt.
j. Rotate CURSOR/DELAY knob and align segmented movable cursor 3 divisions to
left of center vertical line on crt.
k. Press CURSOR SELECT pushbutton and turn CURSOR/DELAY knob to align
next segmented moveable cursor 3 divisions to right of center vertical line on crt. Display
will indicate within Min/Max indications shown in table 5.
HORIZONTAL
Indications
SEC/DIV
Min
Max
s
2.9700 ms
3.0300 ms
500
l. Set: HORIZONTAL MODE pushbutton to B. Display will indicate within Min/Max
HORIZONTAL
Indications
SEC/DIV
Min
Max
s
2.9700 ms
3.0300 ms
500
m. Select: VERTICAL MODE pushbutton, set: CH 2 to off, and set: CH1 to on.
n. Press CURSOR FUNCTION pushbutton and Time pushbutton to OFF.
o. Select: HORIZONTAL MODE A; HORIZONTAL MODE B will be turned off.
9. Vertical Gain Accuracy
NOTE
If TI does not perform as specified, perform adjustments as
indicated in Section IV.
a. Select: TRIGGER MODE and verify NORMAL is set to on; if not, set: NORMAL to
on.