TB 9-6625-2259-24
(16) Repeat technique of (12) above for remaining TI LEVEL settings listed in table
13. Measuring receiver indication will be within limits specified in table 13.
Measuring receiver
Test instrument
indication
LEVEL settings
(dB)
(dBm)
Min
Max
10
7.4
12.6
3
1.4
5.6
(17) Press LEVEL key and adjust INCR-DECR control for 0.0 dBm indication on
measuring receiver.
(18) Using standard tuned level measurement techniques, verify the measuring
receiver indicates within limits specified in first row of table 14.
(19) Repeat technique of (9) and (10) above for remaining TI LEVEL settings listed
Measuring receiver
Test instrument
indication
LEVEL settings
(dB)
(dBm)
Min
Max
0
-2.6
2.6
-10
-12.6
-7.4
-20
-22.6
-17.4
-30
-32.6
-27.4
-40
-42.6
-37.4
-50
-52.6
-47.4
-60
-62.6
-57.4
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.