TB 9-6625-2262-35
(3) Set calibrator output for settings listed in table 5. TI will indicate within
limits specified.
Test instrument indications
Calibrator output settings
Min
Max
A
297.6 A
A
300
302.4
3
mA
A
A
2976
3024
30
mA1
29.76 mA
30.24 mA
300
mA
297.6
mA
302.4
mA
9.5 A2
9.41 A
9.59 A
1Set function switch to dc mA/A.
2Connect amplifier CURRENT OUTPUT HI and LO to TI A and COM.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI mA/A and COM.
(2) Set function switch to ac A.
(3) Set calibrator output for settings listed in table 6. TI will indicate within
limits specified.
Calibrator output settings
Test instrument indications
Current
Min
Max
40 Hz
A
295.3 A
A
300
304.7
1 kHz
A
295.3 A
A
300
304.7
3
mA
40 Hz
A
A
2953
3047
3
mA
1 kHz
A
A
2953
3047
30
mA1
40 Hz
29.53 mA
30.47
mA
30
mA
1 kHz
29.53 mA
30.47
mA
300
mA
40 Hz
295.3 mA
304.7
mA
300
mA
1 kHz
295.3 mA
304.7
mA
9.5 A
40 Hz
9.34 A
9.66
A
2
9.5 A
1 kHz
9.34 A
9.66
A
1Set function switch to ac mA/A.
2Connect amplifier CURRENT OUTPUT HI and LO to TI A and COM.
b. Adjustments. No adjustments can be made.