TB 9-6625-2262-35
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI VΩ and COM.
(2) Set function switch to Ω.
(3) Set TI and calibrator for settings listed in table 7. At each setting, use calibrator
output adjustment controls to set calibrator control display Reading equal to TI indication.
Calibrator control display Error indications will be within limits specified in table 7.
Calibrator
Error
Test instrument
indications
Output
range settings
(%)
settings
Ω1
320 Ω
0.42
190
0.26
3.2 kΩ
1.9 kΩ
0.26
32 kΩ
kΩ
19
kΩ2
320 kΩ
0.26
190
0.26
3.2 MΩ
1.9 MΩ
1.05
32 MΩ
MΩ
19
Set calibrator 2 wire Comp to ON.
1
Set calibrator 2 wire Comp to OFF.
2
(4) Press RANGE pushbutton until nS is displayed.
(5) Set calibrator for a 100 MΩ output. TI will indicate between 9.7 and 10.3 nS.
b. Adjustments. No adjustments will be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.