TB 9-6625-2265-35
a. Performance Check
(1) Connect calibrator OUTPUT terminals to TI V-Ω and COM terminals and set
TI function/range dial to DCV 2.
(2) Set TI function/range dial and calibrator output as specified in table 3. If TI
does not indicate within the specified limits, perform b below.
b. Adjustments. Set TI function/range dial to DCV 200m and calibrator for a 1 0 mV dc
9
output. Adjust VR1 DC 200 mV (fig. 1) for a TI indication of 190.0 mV (R).
Test instrument
Calibrator
Test instrument
function/range dial
output
(DCV)
(V dc)
Min
Max
2
1.9
1.897
1.903
200m
.19
189.7
190.3
200m
-.19
-189.7
-190.3
20
19
18.97
19.03
200
190
189.7
190.3
1000
1000
998
1002