TB 9-6625-2266-35
(4) Repeat technique of (2) and (3) above using settings listed in table 3 below. If
TI does not indicate within limits specified in table 3, perform b below.
Calibrator
Test instrument
Range
Indication limits1
Output
setting
Min
Max
190
mV
200 mV
189.4 (189.7)
190.6
(190.3)
-190
-189.4 (-189.7)
-190.6 (-190.3)
200 mV
mV
1.9 V
2V
1.894 (1.897)
1.906
(1.903)
19
V
20 V
18.94 (18.97)
19.06
(19.03)
190
V
200 V
189.4 (189.7)
190.6
(190.3)
1000
V
1000 V
997
(998)
1003
(1002)
Values in parenthesis apply to model 8020B.
1
b. Adjustments. Set TI range to 200 mV and calibrator for a 190 mV dc output.
Adjust DC CAL R6 (fig. 1) for a TI indication of 190.0 (R).
6