TB 9-6625-2267-35
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and
Common name
Minimum use specifications
model (part number)
CALIBRATOR
Dc voltage range: -190 mV to 1000 V
John Fluke, Model
Voltage:
Accuracy: (%)
5700A/CT (p/o MIS-
190 mV
..............................................013
35947), w/power
-1.9 to 190 V ..............................................0066
amplifier, John Fluke,
1000 V
..............................................0075
Model 5725A (5725A)
Ac voltage:
Range: 190 mV to 1000 V
Accuracy: (%)
Voltage
.030
5.0
.040/20
50
100
190 mV
.151
.071
N/A
.151
.257
1.9 V through
.132
.0541
N/A
.132
.263
190 V
400 V
N/A
.075
N/A
N/A
N/A
1000 V
N/A
.102
.150
N/A
N/A
Dc current range: 190 A to 1.9 A
Current:
Accuracy: (%)
190 A to 190 mA ........................................
.028
1.9 A
..............................................053
Accuracy: (%)
100Ω
..............................................0325
1.0 kΩ
..............................................0275
10, 100, and 1000 kΩ ..................................
.0150
10 MΩ
..............................................0525
6. Preliminary Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration process. Personnel should become familiar with the entire bulletin before
beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Additional maintenance information is contained in the manufacturer's manual for this
TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.