TB 9-6625-2267-35
(8) Set calibrator for a 1.9 V, 50 kHz output.
(9) Press RANGE 2 pushbutton. Adjust Cl9 (fig. 1) until TI indicates 1.9000 5
digits (R).
(10) Repeat (6) through (9) above until no further adjustments are required.
(11) Press RANGE 20 pushbutton.
(12) Set calibrator for a 19 V, 50 kHz output. Adjust C10 (fig. 1) until TI indicates
19.000 10 digits (R).
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI INPUT MA and LO.
(2) Press TI FUNCTION DC MA.
(3) Press TI RANGE pushbuttons and set calibrator output for settings listed in
table 6. TI will
indicate within limits
specified.
Test instrument
Test instrument
RANGE
Calibrator
indications
pushbuttons
output settings
Min
Max
200 A
A
190
189.79
190.21
2
1.9
mA
1.8979
1.9021
20
19
mA
18.979
19.02l
200
190
mA
189.79
190.21
2000 mA
1.9
A
1896
1904
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.