TB 9-6625-2274-50
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
parameters
Performance specifications1
Range: 40 V p-p to 200 V p-p into 1M Ω2
Amplitude calibrator: Dc
and squarewave (chopped
40 V p-p to 5 V p-p into 50 Ω
dc)
(.25% of reading +1 V)
Accuracy:
Variable range: 9.9%
Current amplitude mode
Range: 1 to 100 mA
(.25% of reading +2 A)
Accuracy:
Edge amplitude mode
Range: Low amplitude output into 50 Ω, 20 mV p-p to 1 V p-p
Risetime: ≤1.3 ns
Leading edge aberrations: 2% of square wave V p-p amplitude
Long term flatness: Droop and tilt 0.5% maximum of pulse
amplitude after first 100 ns from leading
edge 50% amplitude point
Frequency: 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, and 1 MHz
Range: High amplitude output into 1 M Ω, 1 V p-p to 100 V p-p3
Risetime: ≤100 ns
Leading edge aberrations: 2%. of square wave V p-p amplitude
Long term flatness: Droop and tilt 0.5% maximum of pulse
amplitude after first 500 ns from leading
edge 50% amplitude point
Frequency: 10 Hz, 100 Hz, 1 kHz, 10 kHz, and 100 kHz
Variable amplitude: ≤9.9% from nominal
Timing markers mode
Range: .5 ns to 5 s
Accuracy:
(.005% of reading +10 ps)
Amplitude: 5 s to 2 ns ≥l V p-p into 50 Ω
1 ns ≥350 mV p-p into 50 Ω
.5 ns ≥100 mV p-p into 50 Ω
Variable marker: 9.9% of marker setting
Trigger output
Trigger amplitude: 1 V minimum into 50 Ω
Trigger NORMAL rate: 5 s to 100 ns; remains at 100 ns for faster rates
Trigger mode RATE10: Divides NORMAL rate by 10
Trigger mode RATE 100: Divides NORMAL rate by 100
Programmable Pulse Head, Tektronix, Type 015-0611-00
Fast edge amplitude
Range: 1. 1 V 5% into 50 Ω
mode
Risetime: <150 ps
Leading edge aberrations: 3% pulse amplitude; not to exceed 40%, p-p
for adjacent peaks
Frequency:
100 Hz, 1 kHz, 10 kHz, and 100 kHz
Variable range: >l0%
1Specifications
are IAW MIS-35947 and may not agree with manufacturer's stated accuracies.
2200
V p-p check not verified due to standards limitation.
3100 V p-p check not verified due to standards limitation.
4Long term flatness not verified due to standards limitation.