TB 9-6625-2275-24
(6) Reconnect equipment as shown in figure 1.
(7) Set calibrator to OPERATE and multimeter to 100 mV range.
(8) Subtract value recorded in (1) above from multimeter indication. Resulting
value will be between 18.9867 mV and 19.0133 mV dc.
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
CALIBRATION PROCESS FOR
FLUKE, MODEL 80J-10
10. Preliminary Instructions
a. The instructions outlined in paragraphs 10 and 11 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
d. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe the safety precautions. REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Set multimeter for dc voltage measurements. Short VOLTS INPUT and then zero
the multimeter.
b. Connect equipment as shown in figure 2 CONNECTION A.