TB 9-6625-2276-24
NOTE
Disregard sign of difference determined in (18) above.
(19) Repeat (18) above for remaining test frequency settings listed in table 3.
PRE-AMP gain for test frequencies listed in table 3 will be greater than 25 dBm.
NOTE
(For information only)
Typical PRE-AMP gain is between 28.5 and 31.5 dBm.
(20) Add highest and lowest PRE-AMP gain recorded in table 3 and then divide sum
by two. Highest and lowest PRE-AMP gain values will be within 1.5 dBm of
calculated average.
(21) Disconnect TI from equipment setup.
(22) Transcribe reference column values recorded in table 3 to reference column in
(24) Adjust signal generator frequency controls to frequency setting listed in table 4
and amplitude controls for a measuring receiver indication equal to reference recorded in
6