TB 9-6625-2277-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Digital Multimeter, Fluke, Models 8010A, 8010A-01, 8010M, and 8012A. The
manufacturers' manuals and/or technical manuals were used as the prime data sources in
compiling these instructions. The equipment being calibrated will be referred to as the TI
(test instrument) throughout this bulletin.
a. Model Variations. Variations among models are described in text and tables.
b. Time and Technique. The time required for this calibration is approximately 1
hour, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are in table 1.
Performance specifications
Test instrument
(% of reading + digits)
parameters
(3 1/2 digit display)
Range: 200 mV, 2V, 20V, 200V, 1000V (1200V)1
Dc voltage
Accuracy: (0.1 + 1)
Ac Voltage
Range: 200 mV, 2 V, 20 V
Accuracy: 45 Hz to l0 kHz
(0.5 + 2)
10 to 20 kHz
(1.0 + 2)
20 to 50 kHz
(5.0 + 3)
Range: 200 V
Accuracy: 45 Hz to 10 kHz
(0.5 + 2)
10 to 20 kHz
(1.0 + 2)
20 to 50 kHz1
(5.0 + 3)
Range: 750 V
Accuracy: 45 Hz to 1 kHz
(0.5 + 2)
Range: 1200 V1
Accuracy: 45 Hz to 10 kHz
(0.5 + 2)
10 to 20 kHz
(1.0 + 2)
See footnotes at end of table.