TB 9-6625-2277-24
Figure 1. Adjustment locations.
a. Performance Check
(1) Connect calibrator OUTPUT HI to TI V/k S (V/k
on model 8010M) and
calibrator OUTPUT LO to TI COMMON.
(2) Press V function pushbutton and AC/DC pushbutton to AC.
(3) Press TI range pushbutton and set calibrator output for settings listed in table 4. If
TI does not indicate within limits specified in table 4, perform b below.
Test instrument
range
Calibrator settings
Test instrument indications
pushbuttons
Voltage
Min
Max
200 mV
190 mV
45
Hz
188.8
191.2
200 mV
190 mV
1
kHz
188.8
191.2
200 mV
190 mV
10
kHz
188.8
191.2
200 mV
190 mV
20
kHz
187.9
192.1