TB 9-6625-2277-24
1Model
8010M.
NOTE
Perform (4) through (9) below for models 8010A and 8010A-01.
and 2 range pushbuttons to select 2 mS range.
(4) Simultaneously press 200
nominal output with 2 wire comp ON.
(5) Set calibrator for 1000
TI will
indicate between 0.997 and 1.003.
(6) Simultaneously press 20 and 200 range pushbuttons to select 20 S range.
(7) Set calibrator for 100 k nominal output. TI will indicate between 9.97 and 10.03.
range pushbuttons to select 200 nS range.
(8) Simultaneously press 2000 and 20M
(9) Set calibrator for 10 M
nominal output. TI will indicate between 98.0 and 102.0.
b. Adjustments. No adjustments can be made.
12. Resistance (Model 8012A only)
a. Performance Check
and calibrator OUTPUT
(1) Connect calibrator OUTPUT HI to TI LO RANGE
LO to TI COMMON.
f
(2) Simultaneously press V and mA function pushbuttons to select LO RANGE
unction and press 2 range pushbutton.
NOTE
Calibrator 2 wire Comp must be set to ON in (3) through (6)
below.
(3) Set calibrator output for 0 ohm. Adjust TI ZERO control for a .000 to .001 TI
indication.
(4) Set calibrator for a nominal 1.9 ohm output. Rotate calibrator knob below EDIT
FIELD pushbutton to adjust calibrator display indication to equal TI indication. If
calibrator err display does not indicate within 1.105%, perform b below.
(5) Press 20 range pushbutton.
output. Rotate calibrator knob below EDIT
(6) Set calibrator for a nominal 19
FIELD pushbutton to adjust calibrator display indication to equal TI indication. If
calibrator err display does not indicate within 0.632%, perform b below.
(7) Connect calibrator output to TI V/k /S and COMMON.
Press k /S
function pushbutton.
(8) Press TI range pushbutton and set calibrator to the nominal resistance outputs
as listed in table 7. At each resistance output, rotate calibrator knob below EDIT FIELD
pushbutton to adjust calibrator display indication to equal TI indication. Calibrator err
display indication will be within limits specified in table 7.