TB 9-6625-2282-24
NOTE
Rotate the night vision detector standard a few degrees at a
time then remove hand to allow reading to stabilize. Continue
until maximum indication is obtained.
(6) Set TI HIGH LOW switch to HIGH and rotate night vision detector standard in
TI collimator port to obtain peak output voltage. Multimeter will indicate between 1.56 V
dc and 4.74 V dc.
(7) Set II, III, OFF switch to III. Multimeter will indicate between 0.703 V and
2.18 V dc.
(8) Set HIGH LOW switch to LOW.
(9) Set II, III, OFF switch to II. Multimeter will indicate between 8.4 mV dc and
16.9 mV dc.
(10) Set II, III, OFF switch to III. Multimeter will indicate between 3.6 mV dc and
7.2 mV dc.
(11) Set II, III, OFF switch to OFF.
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment and reinstall protective cover on TI.
b. Annotate and affix DA label/form in accordance with TB 750-25.