TB 9-6625-2288-35
(22) Adjust calibrator for .03 V ac output. Adjust 30MV HI (fig. 1) for a meter
indication between .0285 and .0315 V ac (ideal .03) (R).
(23) Adjust calibrator for .02 V ac output. Adjust 30MV MID (fig. 1) for a meter
indication between .0185 and .0215 V ac (ideal .02) (R).
(24) Adjust calibrator for .01 V ac output. Adjust 30MV LO (fig. 1) for a meter
indication between .0085 and .0115 V ac (ideal .01) (R).
(25) Repeat (22) through (24) above until no further adjustments are required.
(26) Set TI RANGE-FULL-SCALE switch to .1-VOLTS.
(27) Adjust calibrator for .1V ac output. Adjust .1V HI (fig. 1) for a meter indication
between .095 and .105 V ac (ideal .1) (R).
(28) Adjust calibrator for .06 V ac output. Adjust .1V MID (fig. 1) for a meter
indication between .055 and .065 V ac (ideal .06) (R).
(29) Adjust calibrator for .03 V ac output. Adjust .1V LO (fig. 1) for a meter
indication between .025 and .035 V ac (ideal .03) (R).
(30) Repeat (27) through (29) above until no further adjustments are required.
(31) Set TI RANGE-FULL-SCALE switch to .3-VOLTS.
(32) Adjust calibrator for .3 V ac output. Adjust .3V HI (fig. 1) for a meter indication
between .285 and .315 V ac (ideal .3 V) (R).
(33) Adjust calibrator for .2 V ac output. Adjust .3V MID (fig. 1) for a meter
indication between .185 and .215 V ac (ideal .2) (R).
(34) Adjust calibrator for .1V ac output. Adjust .3V LO (fig. 1) for a meter indication
between .085 and .115 V ac (ideal .1) (R).
(35) Repeat (32) through (34) above until no further adjustments are required.
(36) Set TI RANGE-FULL-SCALE switch to 1 -VOLTS.
(37) Adjust calibrator for 1 V ac output. Adjust 1V HI (fig. 1) for a meter indication
between .950 and 1.05 V ac (ideal 1) (R).
(38) Adjust calibrator for .6 V ac output. Adjust 1V MID (fig. 1) for a meter
indication between .550 and .650 V ac (ideal .6) (R).
(39) Adjust calibrator for .3 V ac output. Adjust 1V LO (fig. 1) for a meter indication
between .250 and .350 V ac (ideal .3) (R).
(40) Repeat (37) through (39) above until no further adjustments are required.
(41) Set TI RANGE-FULL-SCALE switch to 3-VOLTS.
(42) Adjust calibrator for 3 V ac output. Adjust 3V HI (fig. 1) for a meter indication
between 2.85 and 3.15 V ac (ideal 3) (R).
(43) Adjust calibrator for 2 V ac output. Adjust 3V MID (fig. 1) for a meter indication
between 1.85 and 2.15 V ac (ideal 2) (R).
(44) Adjust calibrator for 1 V ac output. Adjust 3V LO (fig. 1) for a meter indication
between .850 and 1.15 V ac (ideal 1) (R).
7