TB 9-6625-2294-24
SECTION
MENU CALIBRATION PROCESS
Preliminary instructions ....................................
Equipment setup.................................................
Power supply .......................................................
CAL 01 horizontal...............................................
CAL 02 vertical ...................................................
CAL 03 triggering ...............................................
CAL 06 vertical transient response ...................
Final procedure ...................................................
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Oscilloscope, Tektronix, Types 2465A and 2467. The manufacturer's manual
was used as the prime data source in compiling these instructions. The equipment being
calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 6
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specification
Range: 2 mV/div to 5 V/div
Vertical gain CH 1 and CH 2
2% (1 M input coupling)
Accuracy:
(1.25% of reading + 0.03 div)
Delta V cursors
Accuracy:
CH 3 and CH 4
Range: 0.1 and 0.5 V/div
10%
Accuracy: